image1 image2

Product Search

Category

Measuring Division
Equipment Division
Injection Division
Early Wildfire Detection System
CS CENTER

If you have any questions, please contact us.

TEL: +82-54-473-2900

DEPARTMENT
Measuring Division
Equipment Division
Injection Division
Early Wildfire Detection System
 
Measuring Division
Temperature and Humidity Chamber
GTPS-1THR4000ES
Origin  : KOREA
Brand  : GTPS
Manufacturer  : NDK
Product Details
Related product
1 SIZE : IN : 2000¡¿ 4000¡¿ 2300 [ W¡¿ D¡¿ Hmm]
        OUT : 2460 X 5856 X 2485  [ W¡¿ D¡¿ Hmm]


2 MATERIAL : IN : SUS304 CP 0.5T
        OUT : SS41 0.5T  COATING  Floor 2T


3 Temp and Humidity
¨ç Temp Working range : 20¡É¢¦+100¡É
¨è Humidity Working range : 30%¢¦98%R.H
¨é Resol. temp. : ¡¾ 0.1¡É 
¨êResol. humid : ¡¾ 0.1% R.H
¨ë Temp Unif ormity : ¡¾ 1.0¡É
¨ì Humidity  Unif ormity : ¡¾ 3.0% R.H
4 HUMIDITY. RANGE :  30~ 95%
     at 20¡É :  30~ 90%
     at 30¡É :  25~ 95% 
     at 40¡É :  25~ 95% 
     at 50¡É :  25~ 95% 
     at 60¡É :  25~ 95% 
     at 70¡É :  20~ 95% 
     at 80¡É :  20~ 95%
5 1) CONTROL: Simultaneous temperature and humidity control touch screen Samwon (temi- 2500) or Nexcon (NEX- 1200)
¨ç RS232. Built-in 485 communication
¨è Real-time graph monitoring function
¨é The humidification method of the chamber is an internal humidification method 
   and the temperature and humidity control of the P.I.D SSR type.
¨ê Temperature sensor PT100Ώ
2) Trouble monitoring due to each drive safety device and separate response of program end signal 
   and trouble signal with multiple buzzers
3) The MAGNETIC RELAY of refrigerator control uses parts with an
   OVER LOAD attached, and the OVER LOAD is variable.
   (EOCR attachment time, AMPERE variable setting)
4) The power source uses AC220V 3ϕ 60HZ or AC380V 3ϕ 60HZ 22kw.
5) Observation Window : 350mm X 450mm
6 Heater
1) Dry bulb heater: SUS316 12.0KW - Made of SUS316 to ensure corrosion resistance.
2) Wet bulb heater: SUS316 6.0KW - Made of SUS316 to ensure corrosion resistance.
All connections are delta connections and use Teflon- and silicone-coated asbestos grade heat-resistant wire.


7 COMPRESSORS
¨ç Semi-closed air-cooled primary: Copeland 3HP
   - Use for low temperature to ensure cooling effect with high noise and vibration performance.
   - Maximize energy saving effect with electronic expansion valve control
   - The temperature of the heat exchanger is monitored in real time with an electronic 
     expansion valve to prevent freezing during long-term operation.


8 Water supply device: automatic water supply
9 Air circulation motor
¨ç BLOWER MOTOR: 1/ 2HP Siemens Motor (Germany)
¨è FAN BLOWER: To achieve even distribution, blow a 100¡¿200¡° SIROCCO FAN along the wall.
    Securing high distribution by dispersing it overall
¨é FAN material: Made of aluminum to prevent rust
10 Energy saving device
Compared to existing equipment and other companies, 30~40% electricity cost savings, 
low heat dissipation, low noise, low vibration, up to 70% energy savings, and humidification water savings.
11 Air circulation method: Forced circulation method, Impeller 200¨ª BLOWER FAN 4EA
12 Safety Devices : Fan over current, temp overheating, water empty, humidity overheating, 
                    Ref rigerating over current, ref rigerating high presser/ low presser 
                    Emergency switch
- It is designed as a system that displays the abnormal part on the controller and transmits a warning sound
  to the surrounding area when an abnormality occurs during operation with safety devices at each drive.
- Emergency switch allows Emergency stop while driving
- Noise can be muted with a separate BUZZER SWITCH










Product Details
Related product
There are no products associated with this product.
Product Q&A
No. Subject Name Date Remarks
No Product Inquiry created.
  Contact Us