¢Â UV-3600i Plus Mainbody
1) Optical system : Double beam metering system
2) Light source : Halogen lamp, Deuterium lamp
(Includes automatic optical axis adjustment mechanism)
3) Light source switching : Automatic switching
(Optional switching wavelength: 282-393 nm, Settable in 0.01 nm steps)
4) Spectroscope : Double monochromator
5) Detector : UV/VIS£ºPhotomultiplier tube
NIR£ºInGaAs Photodiode
CooledPbs Photoconductive element"
6) Wavelength range : 185–3300 nm
7) Wavelength accuracy : UV/VIS¡¾ 0.2 nm, NIR¡¾ 0.8 nm
8) Wavelength repeatability : Within UV/VIS¡¾ 0.08 nm, Within NIR¡¾ 0.32 nm¡¡
9) Spectral bandwidth : UV/VIS 0.1, 0.2, 0.5, 1, 2, £³, 5, 8 nm
NIR 0.2, 0.5, 1, 2, 3 , 5, 8,12 ,20, 32 nm
10) Stray light : 0.00008 £¥ or less (220 nm, NaI)
0.00005 £¥ or less (340, NaNO2)
0.0005 £¥ or less (1420 nm, H2O)
0.005 £¥ or less (2365 nm, CHCl3)
11) Photometric range : Absorbance£º −6 to 6 Abs
12) Photometric accuracy : ¡¾ 0.002 Abs (0.5 Abs), ¡¾ 0.003 Abs (1 Abs)
Test using NIST930
13) Photometric repeatability : 0.0008 Abs (0~0.5 Abs), 0.0016 Abs (0.5¢¦1 Abs)
Max. Deviation of 5 measurements per second
14) Wavelength scanning speed : UV/VIS£ºMax.Approx.4500nm/min
Near infrared PMT/InGaAs region£ºMax. approx. 9000nm/min
NIR£ºMax.approx. 70000nm/min
CooledPbs region£ºMax.4000nm/min
15) Wavelength transfer rate : UV/VIS£ºMax.approx. 18000nm/min,
NIR£ºMax.approx. 70000nm/min
16) Noise level : 0.00005Abs or less(RMS 500nm),
0.00008Abs or less(RMS 900nm),
0.00003Abs or less(RMS 1500nm)
SBW2nm
17) Baseline stability : Within 0.0002 Abs/h(500 nm), 2 H after light source turned ON
18) Baseline flatness : ¡¾ 0.004 Abs (185–200 nm),
¡¾ 0.001 Abs (200-3000 nm),
¡¾ 0.005 Abs (3000-3300 nm)
19) Dimensions: W1020 ¡¿ D660 ¡¿ H286 mm
20) Weight : 96 kg